Computer control 30T cement pressure tester (fault phenomenon, cause and elimination method)
Computer control 30T cement pressure tester (fault phenomenon, cause and elimination method)
1, Pipe fittings at the leakage of oil
One of the reasons is that the sealing
connection is loose, in which case the components should be properly assembled
and tightened, the second reason is the seal damage, in this case the seal
should be replaced with seals such as 0 seals, combination washers or clamp,
when replacing seals with kerosene cleaning seals and fittings to prevent
contaminants from entering the hydraulic system, damaging hydraulic components
(Such as commutator valve, proportional valve).
2. Test Power Overshoot
The most important reason for the test
force overshoot is that the control parameters are not accurate, first of all,
we should consider whether the test force rise speed is correct, if not correct
should modify the test force rise speed, and then see if the PID has been
adjusted, if there is no should adjust the value of the PID.
3, The pressure does not go
Cause the pressure does not go the reason
should first consider whether the commutator valve is stuck in the oil delivery
position, this situation should be cleaned or replaced the commutator valve,
and then see if the motor is reversed, if the power cord is any two FireWire
position interchange, and finally should check whether the manual valve is in
the oil return position, if the manual valve is placed in the hold position
4.Piston does not fall back
Check whether the commutator valve is
stuck, if it is stuck should clean or replace the commutator valve, and then
see if there is no signal to the sensor, if there should be no inspection of
the electromagnetic commutator valve related to the reliability of the line
connection and A/D card has no output.